CAN DIEN TU | Shoping
Home page About us FAQs Services Certificate Contact us Shopping cart Payment

Please be informed that our Dong Nhan company will be close for Lunar New Year holidays from 11 Feb. 2010 and resume on 23 Feb. 2010

 Việt Nam  English
PRODUCTS
SCALE FISHERIES
PACKING BALANCE
SCALE CARS
Check Warranty
FORUM


Moisture analyzer model : MS-70 - MX-50 - ML-50 - MF-50 AND

Delivery time : available

 The AND Weighing MF, MS and MX series moisture analyzers are ideal for food, pharmaceutical, chemical processing, or any moisture analysis application, particularly those needing GLP/GMP/ISO compliance. This one of a kind drying system, combined with an advanced heat protection mechanism for the weigh cell, provides superior accuracy and repeatability

Download User manual : HDSD can MX-50 AND

 Features :

  • Fast and uniform heating with halogen lamp and innovative SRA technology
    Straight halogen lamp and uniquely designed SRA (Secondary Radiation Assist) filter gives shorter measurement time, thanks to fast and uniform heating.

  • High Repeatability
    With SHS (Super Hybrid Sensor) featured as the weight sensor, ultra accurate moisture content determination is possible based on high precision weighing of even a small sample

  • High moisture content measurement
    MS-70 measures the moisture content at 0.001% resolution suitable for low moisture content samples as well as Karl Fischer method, yet requires no special knowledge or training and produces no harmful waste

  • Standard WinCT-Moisture (for MS & MX) for real-time graph displaying
    WinCT-Moisture is an original software application designed to display a graph of moisture content rate change while measuring with a connected PC

  • Sodium Tartrate Dihydrate comes a s standard for accuracy checking
    Sodium Tartrate Dihydrate is a chemical material that has stable moisture content of 15.66% and thus is best for use for accuracy check to maintain the reference value of the analyzer

  • Calibration of the heater temperature (for MS & MX)
    With the temperature calibrator (optional), calibration result can be output in the format that conforms to GLP, GMP, ISO

  • Memory Function - According to sample up to 20 suitable measurement conditions can be stored and recalled, which saves time and prevents the user from making a mistake when setting (10 for MF)

  • Five Measurement Programs - Five choices of measurement programs: Standard, Automatic, Quick, Timer and Manual Mode are provided:

  • Standard Mode: just measurement accuracy: HI, MID or LO needs to be set

  • Automatic Mode: Ends measurment when moisture content changes at a rate less than the set rate

  • Quick Mode: Begins heating samples at 200? C for 3 minutes, then is the same as Automatic Mode

  • Timer Mode: Continues measurement for a set duration of time (1-60 minutes by 1 minute, 60-480 min by 5 min)

  • Manual Mode: STOP button should be pressed to end measurement (max heating time = 480 minutes)

  • Selectable Heating Mode
    Choose the heating mode from standard, quick, step and ramp heating modes for the most suitable measurement (ML has Standard and Quick heating modes only)

  • Clear and easy to see, large VFD display
    Measurement, setting value, change in moisture content, action status, data number and other important information are clearly displayed

  • Easy handling of the unit
    Ergonomically designed pan handle eliminates mishaps such as burns when moving a hot sample pan into or out of the unit from either side. Wing handle for easy opening and closing of the heater cover

  • Low maintenance cost assured
    The halogen lamp is user replaceable without unit downtime with protective chamber for easy cleaning (lamp life is 5000 hours)

  • Progress window for heating check
    Heating process can be checked through the translucent window

  • Self Check function
    Deflect check function is available with temperature control

  • Quick Reference card
    A convenient operating guide is installed at the bottom of the analyzer

  • Standard RS-232C
    Bi-directional communication with a PC or connects directly to a printer

  • Conformity to GLP, GMP, GCP and ISO with date/time, ID, calibration data and check record outputs

  • Data output for daily record management

..:: 12 MONTHS WARRANTY ::..

 

Specifications : 

Model

MS-70

MX-50

MF-50

ML-50

Maximum Sample Weight Capacity

71g

51g

Weight Resolution

0.001g

0.001g

0.002g

0.005g

Moisture Content Display (%)

0.001/0.01/0.1

0.01/0.1

0.05/0.1/1

0.01/0.1

Moisture Content Accuracy-over 1g(%)

0.05

0.10

0.20

0.5

Moisture Content Accuracy-over 5g(%)

0.01

0.02

0.05

0.1

Memory of Measurement Programs

 

20 sets

10 sets

 

Data Memory Function

 

100

50

 

Communication Software

WinCT standard

WinCT Moisture Standard

WinCT standard

Heating Technology

Halogen lamp (straight type: 400 watt max, 5000 hours)

Drying Temperature

30-200?C

500C 2000C (10 C increments)

Measurement Programs

Standard Mode, Automatic Mode, Quick Mode, Timer Mode, Manual Mode

Display Type

Large VFD

Interface

RS-232C standard

Temperature

50C ~ 400C (140F ~ 1040F)

GLP/GMP/ISO

Available

Self Check Function

Standard

Sample Pan Size

2.16 inch

Power

AC 100V to 120V or AC 200V to 240V, 50/60 Hz, approximately 400 W

Dimensions

5.46 W x 8.1 D x 4.4 H inch
(215 W x 320 D x 173 H)

Price(USD)

call

call

call

call